Spectrometer SM 9000

The new Spectrometer SM 9000 was developed for high-resolution, fiber optic spectroscopy applications in the UV/VIS to NIR range.

  • Product Name: Spectrometer SM 9000
  • Model:SM 9000
  • Maunufactor:PSI
  • Country:CZ

The new Spectrometer SM 9000 was developed for high-resolution, fiber optic spectroscopy applications in the UV/VIS to NIR range. It can be used as a stand-alone instrument or as a substantial enhancement of other PSI instruments to analyze spectral composition of irradiance. Ultra-high sensitivity of the instrument allows measuring fluorescence emission spectra of individual cells when combined with µ-FluorCam. The spectra can be measured in the range from UV (200 nm) to NIR (980 nm) with FWHM bandwidth of 3.5 nm. Fast recording can capture as many as 100 spectra per second with data resolution of 16 bits.


KEY FEATURES:

  • Ultra-high sensitivity.

  • Suitable for microspectrophotometry.

  • FWHM bandwidth of 3.5 nm.

  • Small-footprint, modular, robust, and thermally stable.

  • Very low heat generation.

  • 10 µs - 10 ms duration of flashes.

  • Variable exposure time, from 1 millisecond to minutes.

  • Kinetic spectra (max. 100 per second).





Optical Entrance

diameter 0.5 NA = 0.22 mounted in SMA-coupling, dismountable

Entrance slit

70 µm x 1400 µm (optical entrance)

Grating

Flat-field correction

Spectral range

200 nm - 980 nm

Wavelength accuracy absolute

< 0.5 nm

Reproducibility

< 0.1 nm

Temperature - induced drift

<0.01 nm/K

Spectral distance of pixel

Δλ pixel ~ 0.8 nm

(FWHM: Full width half maximum)

Δλ < 3 ...4 nm (UV-NIR Version)

Straylight

0.1 % measured at 340 nm with deuterium lamp (transmission of NaNO2 solution, 50 g/l, 1 cm)

CCD Array

Thermoelectrically cooled Hamamatsu S 7031

Number of pixels

1044 x 64

Dimensions of pixels

24 x 24 mm2

System data

16 Bit AD conversion

Noise

2...4 count standard deviation