SwisensPoleno Jupiter is the latest generation of optical particle measurement systems for real-time measurement and monitoring of bioaerosols. The real-time bioaerosol monitor SwisensPoleno Jupiter combines the latest measurement methods with artificial intelligence and transparent data evaluation to create a reliable measurement system for automatic particle identification. This includes pollen, spores, other bioaerosols and solid particles in the ambient air.
Designed for discovering the particle microcosm and pioneering autonomous measurement and monitoring of bioaerosols.
SwisensPoleno Jupiter and its measurement methods:
holographic images
light scattering
polarisation and
UV-induced fluorescence intensity & lifetime
The most important specifications at a glance:
particle classes within 0.5 – 300µm
air sampling volume 40l/min
sigma-2 geometry sample inlet
integrated particle concentrator
Further benefits of SwisensPoleno Jupiter
non-invasive measurement method
immediate verification of identification results
remote access to data
fully remote operation, configuration and updates
integrated self-cleaning function
customized weatherproof housing included
no consumables
Environmental conditions Outdoor -20°C to 50°C, 0 to 1 RH; non-corrosive environment (please consult us if you need to use near the sea)
Inlet port σ2 geometry (VDI 2119 Ambient Air Measurement, 2.5μm particle sampling) Mesh size: 1mm
Particle measurement range 0.5µm-300µm
Particle velocity 0.5m/s
Concentration factor 1000 (Particle diameter: 10µm to 300µm)
Measurement Rate Maximum 30000 particles/m3 ( Particle diameter: 10μm to 300μm)
Maximum 1,000,000 particles/m3 (Particle diameter: 2μm)
Holographic Measurement Light source wavelength: 450nm, 685nm
Cameras: USB3 cameras, 2 at 90° to each other
Fluorescence measurement Light source wavelengths: 280nm, 365nm and 405nm
Detectors: SiPM (Silicon Photomultiplier), 5 pcs.
Scattered and polarised light measurements Laser wavelength: 405nm
Receivers: 2, horizontally and vertically polarised
Communication Ethernet (if not using an integrated mobile router), 4G
External interfaces 24VDC powered, Ethernet, HDMI, RS-232/RS-485, USB 2.0 (mouse and keyboard), USB 3.0 (expansion devices), digital port, analogue port
Networking Functions Support
Algorithm Extension Support
Weather Sensor Expansion Supported
IP Rating IP65 (main chassis), IP54 (air conditioning)
Dimensions 63 x 73 x 150 cm3 W x D x H (including protective case, sample inlet and mounting bracket)
Weight 134 kg
Power Supply 100-240 VAC, 50/60Hz, 750W peak (incl. industrial control unit and air conditioning)
Outdoor and Indoor Applications
SwisensPoleno Jupiter can be operated in the field as well as in the laboratory and indoors. The installation and removal of the measurement system is done with a few
simple steps. SwisensPoleno Jupiter can be configured accordingly for use in various applications:
• Longterm outdoor monitoring
• Targeted measurements under laboratory conditions
• Reference measurements & algorithm development
• Measurement of pollutants under protective equipment like an aerosol chamber
• Indoor air quality measurement
SwisensAtomizer
The hardware add-ons such as the Swisens-Atomizer enable direct nebulization and measurement of dry particle samples. A HEPA filter plug at the air outlet of the measuring system provides filtering of the measured air volume. Both can be quickly and easily mounted or removed as desired.