:: Product Information

CX402-VFCxxx

Data Loggers

Highlighted Features:


  • Complies with all CDC and Vaccine for Children Program Requirements

  • Easy-to-read display with current and daily min/max temperatures

  • Audible and visual alarms (remote alarms require the InTemp Gateway app)

  • Bluetooth transmission of data and alarm states

  • Logger set up and download of data in seconds – without a PC or cable

  • +/- 0.5C accuracy for reliable temperature readings

  • Battery-powered to guard against local power loss

  • Several Glycol bottle sizes with probe and cable length options

  • Includes 3 Year ISO 17025 compliant calibration 


Specifications


Logger


Radio Power                                 1 mW (0 dBm)

Transmission Range                     Approximately 30.5 m (100 ft) line-of-sight

Wireless Data Standard                Bluetooth Low Energy (Bluetooth Smart)

Logger Operating Range              -30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)

Logging Rate                                1 second to 18 hours

Time Accuracy                              ±1 minute per month at 25°C (77°F)

Battery Type                                 Two AAA 1.5 V alkaline or lithium batteries, user replaceable

Battery Life                                   1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation,                                                                 numerous audible alarms, and paging all impact battery life.

Memory                                         128 KB (84,650 measurements, maximum)

Full Memory Download Time         Approximately 60 seconds; may take longer the farther the device is from the logger


LCD                                              LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures                                          outside this range

Dimensions                                   9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)

Weight                                          90.2 g (3.18 oz)

Environmental Rating                   IP54

CE                                                The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).


 External Temperature Probe


Range                                           -40° to 100°C (-40° to 212°F)

Accuracy                                       ±1.0°C from -40° to -22°C (±1.8°F from -40° to -8°F)

                                                     ±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F)

                                                     ±1.0°C from 50° to 100°C (±1.8°F from 122° to 212°F)

Resolution                                    0.024°C at 25°C (0.04°F at 77°F)

Drift                                              <0.1°C (0.18°F) per year

Response Time                           With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s

                                                   With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s

                                                   With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s

Cable Length                              2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable

NIST Calibration                          Single point NIST calibration, probe only

Probe Dimensions                       Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter


Ambient Temperature Sensor


Range                               -30° to 70°C (-22° to 158°F)

Accuracy                            ±1.0°C from -30° to -22°C (±1.8°F from -22° to -8°F)

                                          ±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F)

                                          ±1.0°C from 50° to 70°C (±1.8°F from 122° to 158°F)

Resolution                         0.024°C at 25°C (0.04°F at 77°F)

Drift                                   <0.1°C (0.18°F) per year


Response Time                 15 minutes typical to 90% in air moving 1 m/s